A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Affected products
The NVD has not attached product data to this record yet; the description names what is affected.
Weakness type
References
- www.cisa.gov /news-events/ics-medical-advisories/icsma-24-319-01
Ingati scans your site every week and tells you when a CVE like this one affects something it found there.