A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specialized equipment and knowledge may be able to configure or disable the pump. No electronic protected health information (ePHI), protected health information (PHI) or personally identifiable information (PII) is stored in the pump.
Affected products
From the NVD's configuration data (CPE), vendor first.
- Bd bodyguard 999-603 firmware, bodyguard 999-603, bodyguard duo 999-903 firmware, bodyguard duo 999-903, bodyguard epidural 999-683 firmware, bodyguard epidural 999-683, bodyguard pain manager 999-803 firmware, bodyguard pain manager 999-803, bodyguard t 999-103 firmware, bodyguard t 999-103, bodyguard 323 colorvision firmware, bodyguard 323 colorvision, bodyguard 121 twins firmware, bodyguard 121 twins
Weakness type
References
- www.bd.com /en-us/about-bd/cybersecurity/bulletin/bd-bodyguard-pumps-r…