A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.
Affected products
From the NVD's configuration data (CPE), vendor first.
- Alivecor kardiamobile firmware, kardiamobile, kardiamobile 6l firmware, kardiamobile 6l, kardiamobile card firmware, kardiamobile card
Weakness type
References
- www.cisa.gov /uscert/ics/advisories/icsma-22-298-01