A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
Affected products
The NVD has not attached product data to this record yet; the description names what is affected.
Weakness type
References
- static.draeger.com /security/download/2021-05-04-PSA-21-124-1-Security-Advisor…
- www.vulncheck.com /advisories/drager-atlan-a350-dos-via-medibus-interface