A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
Innokas Yhtymä Oy Vital Signs Monitor VC150 prior to Version 1.7.15 HL7 v2.x injection vulnerabilities exist in the affected products that allow physically proximate attackers with a connected barcode reader to inject HL7 v2.x segments into specific HL7 v2.x messages via multiple expected parameters.
Affected products
From the NVD's configuration data (CPE), vendor first.
- Innokasmedical vital signs monitor vc150 firmware, vital signs monitor vc150
Weakness type
References
- us-cert.cisa.gov /ics/advisories/icsma-21-007-01