A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
Medtronic MyCareLink Smart 25000 contains
an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
Affected products
From the NVD's configuration data (CPE), vendor first.
- Medtronic mycarelink smart model 25000 firmware, mycarelink smart model 25000
Weakness type
References
- global.medtronic.com /xg-en/product-security/security-bulletins/mycarelink-smart…
- www.cisa.gov /news-events/ics-medical-advisories/icsma-20-345-01
- us-cert.cisa.gov /ics/advisories/icsma-20-345-01
Ingati scans your site every week and tells you when a CVE like this one affects something it found there.