A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Affected products
The NVD has not attached product data to this record yet; the description names what is affected.
Weakness type
References
- static.draeger.com /security/download/2020-06-PSA-20-169-01-Security-Advisory-…
- www.vulncheck.com /advisories/dr-ger-perseus-a500-dos-via-medibus-interface
Ingati scans your site every week and tells you when a CVE like this one affects something it found there.
Start scanning