Vulnerability

CVE-2018-8868

mediumCVSS 3.1 6.2Published 3 Jul 2018

A plain-language summary of this CVE is being written; the description below is the NVD's own text.

Description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Affected products

From the NVD's configuration data (CPE), vendor first.

  • Medtronic 24950 mycarelink monitor firmware, 24950 mycarelink monitor, 24952 mycarelink monitor firmware, 24952 mycarelink monitor