A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
Philips ISCV application prior to version 2.3.0 has an insufficient session expiration vulnerability where an attacker could reuse the session of a previously logged in user. This vulnerability exists when using ISCV together with an Electronic Medical Record (EMR) system, where ISCV is in KIOSK mode for multiple users and using Windows authentication. This may allow an attacker to gain unauthorized access to patient health information and potentially modify this information.
Affected products
From the NVD's configuration data (CPE), vendor first.
- Philips intellispace cardiovascular
Weakness type
References
- www.securityfocus.com /bid/102847
- ics-cert.us-cert.gov /advisories/ICSMA-18-025-01
- www.usa.philips.com /healthcare/about/customer-support/product-security
Ingati scans your site every week and tells you when a CVE like this one affects something it found there.