Vulnerability

CVE-2018-19010

mediumCVSS 3.0 6.5Published 28 Jan 2019

A plain-language summary of this CVE is being written; the description below is the NVD's own text.

Description

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.

Affected products

From the NVD's configuration data (CPE), vendor first.

  • Draeger kappa firmware, kappa, infinity explorer c700 firmware, infinity explorer c700, delta xl firmware, delta xl, infinity delta firmware, infinity delta

Weakness type