A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
Medtronic MiniMed MMT
devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Affected products
From the NVD's configuration data (CPE), vendor first.
- Medtronicdiabetes 508 minimed insulin pump firmware, 508 minimed insulin pump, 522 paradigm real-time firmware, 522 paradigm real-time, 722 paradigm real-time firmware, 722 paradigm real-time, 523 paradigm revel firmware, 523 paradigm revel, 723 paradigm revel firmware, 723 paradigm revel, 523k paradigm revel firmware, 523k paradigm revel, 723k paradigm revel firmware, 723k paradigm revel, 551 minimed 530g firmware, 551 minimed 530g, 751 minimed 530g firmware, 751 minimed 530g
Weakness type
References
- www.securityfocus.com /bid/105044
- global.medtronic.com /xg-en/product-security/security-bulletins/minimed.html
- ics-cert.us-cert.gov /advisories/ICSMA-18-219-02
Ingati scans your site every week and tells you when a CVE like this one affects something it found there.
Start scanning