Vulnerability

CVE-2017-12725

mediumCVSS 3.0 5.6Published 15 Feb 2018

A plain-language summary of this CVE is being written; the description below is the NVD's own text.

Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Affected products

From the NVD's configuration data (CPE), vendor first.

  • Smiths Medical medfusion 4000 wireless syringe infusion pump

Weakness type

References

Ingati scans your site every week and tells you when a CVE like this one affects something it found there.

Start scanning