A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Affected products
From the NVD's configuration data (CPE), vendor first.
- Pfizer lifecare pca infusion system firmware, lifecare pca infusion system
Weakness type
References
- ics-cert.us-cert.gov /advisories/ICSA-15-125-01