A plain-language summary of this CVE is being written; the description below is the NVD's own text.
Description
GE Healthcare Precision THUNIS-800+ has a default password of (1) 1973 for the factory default System Utilities menu, (2) TH8740 for installation using TH8740_122_Setup.exe, (3) hrml for "Setup and Activation" using DSASetup, and (4) an empty string for Shutter Configuration, which has unspecified impact and attack vectors. NOTE: since these passwords appear to be used to access functionality during installation, this issue might not cross privilege boundaries and might not be a vulnerability.
Affected products
From the NVD's configuration data (CPE), vendor first.
- Gehealthcare precision thunis-800\+
Weakness type
References
- apps.gehealthcare.com /servlet/ClientServlet/5458232-1EN+r4.pdf
- www.forbes.com /sites/thomasbrewster/2015/07/10/vulnerable-breasts/
- ics-cert.us-cert.gov /advisories/ICSMA-18-037-02
- twitter.com /digitalbond/status/619250429751222277
Ingati scans your site every week and tells you when a CVE like this one affects something it found there.