Vulnerability

CVE-2014-5433

criticalCVSS 3.0 9.8Published 26 Mar 2019

A plain-language summary of this CVE is being written; the description below is the NVD's own text.

Description

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Affected products

From the NVD's configuration data (CPE), vendor first.

  • Baxter sigma spectrum infusion system firmware, sigma spectrum infusion system, wireless battery module

Weakness type

References

Ingati scans your site every week and tells you when a CVE like this one affects something it found there.

Start scanning