Vulnerability

CVE-2014-5406

highCVSS 2.0 9.3Published 6 Jul 2015

A plain-language summary of this CVE is being written; the description below is the NVD's own text.

Description

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

Affected products

From the NVD's configuration data (CPE), vendor first.

  • Hospira lifecare pcainfusion firmware, lifecare pca3, lifecare pca5

Weakness type

References